Products :
 
  Ultrasonic scalers
  Led lights
  Air polishing Systems
  Dental Units
  Instruments
  Alginates
  Heraeus Kulzer
  Impression paste
  Autoclave
  Air Compressor
  Villa, Italy
  Kdriller
  Impression materials
  Restoratives
 
 
Click to view our After
Sales Service >>
 
Autoclave :
<<back
 
SAFETY DEVICES
Safety Valve- It opens when the pressure reaches from 2.7 to 3 Kgf/cm or from 265 to 294 kPa

Anti-vacuum Valve - It works the same way if the safety valve, relieving the pressure inside the chamber in case of vacuum build up.

Door Gasket - In case the pressure goes beyond 3 Kgf/cm or 294 kPa, the door gasket will detach from the edge of the door making a loud noise.

Fuse-It’s purpose is to protect the building electrical wiring against peaks of energy. The fuse used is the 20 Glass. AGLF-Quick Action 10A/250V for 127V and 6A/250V for 220V.

Thermostat- Internal safety device of the equipment. Its function is to limit the excessive heating of the chamber during the sterilization cycles or in case of a circuit board malfunction.

Electronic Power Control System - Internal system of the equipment which monitors the temperature and the pressure of the autoclave throughout the operation.

Electronic Temperature X Pressure Data Crossing System - Internal System of the equipment which will check the cycle as soon as the autoclave reaches 1070 C, in case any problem is detected while reading the pressure in the chamber or if the autoclave pressure exceeds the safety limit, the cycle will be also cancelled automatically.

 
Mode Heating Time Sterilization Temperature and Pressure Sterilization Time Drying Time
134 C 10 to 30 min 134 C / 216 kPa (2.2 kg/cm) 10 min 30 min
121 C 10 to 30 min 121 C / 118 kPa (1.2 kg/cm) 22 min 30 min
Extra Drying (---------) (---------) (---------) 30 min
    Drying Temperature - 950 C / 2030 F    
 
TECHNICAL SPECIFICATIONS
Technical Data Quadra 12 MecKlave
CertificationsCapacity……………………………………
Weight……………………………………
Weight Per Support Area (N/m2)……………………
Overall Clearance……………………………
Clearance required for the movement of the door………
Chamber Internal Dimension (W x D)…………………
Autoclave External Dimension (W x H x D)………
Voltage……………………………
Frequency……………………………
Power………………………………
Power Consumption………………………………
Maximum And Minimum Steam Pressure……………
Maximum And Minimum Steam Pressure……………
Temperature Of Drained Water………………………
Total heat in joules transmitted in one hour………
Proper working temperature range……………………
Proper working altitude………………………………
Quadra 12 MecKlaveThis equipment is manufactured by Cristofoli, company which Quali Management System is certified and in accordance with the ISO 9001:2000, ISO 1385:2003, GMP- Good Manufacturing Practices and ISO 14001:2004 - Environmental Management standards.12 litersAluminum - 18kg (including support and trays)Stainless Steel - 19.8 kg (including support and trays)42.9 N/m210 cm for each side of the autoclave34 cm22 x 33 cm33.5 x 33 x 48.5 cm127 or 220V Ac (Europe - 230V Ac)50/60 Hz1200 Watts285 Watts Per Cycle0 To 4 Kgf/cm21000c To 1520C1000c771 kj150C To 400Cup to 2500 m.
 
 
  News & Events :
 
To emerge and be recognized as a Domain Leader, which not only is rendering services , and selling high Quality products which meets the highest Clinical Standards, but also to provide, cutting edge technology, and effective After Sales Service.
MORE>>
  Exclusive :
 
Exclusive distributor of KaVo in India
Exclusive distributor of Cavex in India
Exclusive Distributor for Kdriller, Brazil and also exclusive distributor for Schulz, Brazil
Mectron brings the quality of Heraeus back to India
Mectron is happy to announce the representation of Villa, Italy in India and this should lead to the communication attached.
MORE>>
 
  Copyright 2008 Mectron Dental (India) P.Ltd   Sitemap | Legal Disclaimer   Design & Developed by MindSync Interactive